ELISA Kit for Tryptase (TPS)

Catalog Number: USC-SEB070HU
Article Name: ELISA Kit for Tryptase (TPS)
Biozol Catalog Number: USC-SEB070HU
Supplier Catalog Number: SEB070Hu
Alternative Catalog Number: USC-SEB070HU-48,USC-SEB070HU-96
Manufacturer: Cloud-Clone
Category: Kits/Assays
Species Reactivity: Human
Alternative Names: TPSAB1, Tryptase Alpha/Beta 1, Tryptase Alpha II, Tryptase Beta I, Tryptase-I, Tryptase-II, Tryptase-III
48T, 96T, 96T*5, 96T*10, 96T*100

Specificity

This assay has high sensitivity and excellent specificity for detection of Tryptase (TPS).
No significant cross-reactivity or interference between Tryptase (TPS) and analogues was observed.

Recovery

Matrices listed below were spiked with certain level of recombinant Tryptase (TPS) and the recovery rates were calculated by comparing the measured value to the expected amount of Tryptase (TPS) in samples.

Matrix Recovery range (%) Average(%)
serum(n=5) 92-99 96
EDTA plasma(n=5) 82-93 86
heparin plasma(n=5) 97-105 102

Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Tryptase (TPS) were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Tryptase (TPS) were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Tryptase (TPS) and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Sample 1:2 1:4 1:8 1:16
serum(n=5) 98-105% 81-105% 92-102% 80-90%
EDTA plasma(n=5) 91-103% 87-101% 85-103% 80-101%
heparin plasma(n=5) 85-103% 87-96% 90-104% 98-105%

Stability

The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

Reagents and materials provided

Reagents Quantity Reagents Quantity
Pre-coated, ready to use 96-well strip plate 1 Plate sealer for 96 wells 4
Standard 2 Standard Diluent 1×20mL
Detection Reagent A 1×120µL Assay Diluent A 1×12mL
Detection Reagent B 1×120µL Assay Diluent B 1×12mL
TMB Substrate 1×9mL Stop Solution 1×6mL
Wash Buffer (30 × concentrate) 1×20mL Instruction manual 1

Assay procedure summary

1. Prepare all reagents, samples and standards;
2. Add 100µL standard or sample to each well. Incubate 1 hours at 37°C;
3. Aspirate and add 100µL prepared Detection Reagent A. Incubate 1 hour at 37°C;
4. Aspirate and wash 3 times;
5. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
6. Aspirate and wash 5 times;
7. Add 90µL Substrate Solution. Incubate 10-20 minutes at 37°C;
8. Add 50µL Stop Solution. Read at 450nm immediately.

Range: 7.8-500pg/mL
Sensitivity: The minimum detectable dose of this kit is typically less than 2.8pg/mL
UniProt: Q15661
Samples: serum, plasma and other biological fluids
Application Notes: Enzyme-linked immunosorbent assay for Antigen Detection.
Typical Standard Curve USC-SEB070HU